SR EN ISO 80369-7:2017

Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications (ISO 80369-7:2016, Corrected version 2016-12-01)

ISO 80369-7:2016 specifies dimensions and requirements for the design and functional performance of small-bore connectors intended to be used for connections in intravascular applications or hypodermic connections in hypodermic applications of medical devices and accessories. EXAMPLES Hypodermic syringes and needles or intravascular (IV) cannulae with male and female luer slip connectors and luer lock connectors. NOTE 1 Hypodermic use includes percutaneous infusion and injection as well as pressurizing and depressurizing the retention mechanisms (e.g. balloon) used to hold invasive medical devices in place and endoscopic devices. NOTE 2 The luer connector was originally designed for use at pressures up to 300 kPa. ISO 80369-7:2016 does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in particular International Standards for specific medical devices or accessories. This part of ISO 80369 does not specify requirements for the following small-bore connectors, which are specified in other International Standard

50.62

Status : Withdrawn
Approval date : 10/31/2017
Withdrawal date : 7/30/2021
Number of pages : 57
ICS : 11.040.25 Syringes, needles and catheters
Technical Committee : 377 - Medical devices

Relations with other standards: