SR EN IEC 60522-1:2021
Medical electrical equipment - Diagnostic X-rays - Part 1: Determination of quality equivalent filtration and permanent filtration
IEC 60522-1:2020 applies to X-RAY TUBE ASSEMBLIES and to FILTERING MATERIAL, in medical diagnostic applications up to a HIGH VOLTAGE of 150 kV. For HIGH VOLTAGES greater than 50 kV, this document applies to X-RAY TUBE ASSEMBLIES with tungsten or tungsten-alloy TARGETS only. IEC 60522-1:2021 defines the concept of PERMANENT FILTRATION of X-RAY TUBE ASSEMBLIES, and it defines the term FILTERING MATERIAL. Methods are given to determine the PERMANENT FILTRATION of an X-RAY TUBE ASSEMBLY and for determining the QUALITY EQUIVALENT FILTRATION of FILTERING MATERIALS. It contains requirements for statements of compliance of X-RAY TUBE ASSEMBLIES in ACCOMPANYING DOCUMENTS and for markings on X-RAY TUBE ASSEMBLIES, and for indications and statements of compliance of FILTERING MATERIAL. IEC 60522-1:2020 cancels and replaces the second edition of IEC 60522 published in 1999. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the IEC 60522:1999: The scope of the IEC 60522-1 has been changed with respect to se
Status :
Valid
Approval date : 3/31/2021
Number of pages : 46
ICS : 11.040.50 Radiographiic equipment
Technical Committee : 22 - Electrical equipment in medical practice
- Replaces SR EN 60522:2004
RO