This International Standard specifies the requirements, in addition to the resistance versus temperature relationship, for both industrial platinum resistance thermometers (later referred to as "thermometers") and industrial platinum resistance temperature sensors (later referred to as "platinum resistors") whose electrical resistance is derived from defined functions of temperature. Values of temperature in this document are in terms of the International Temperature Scale of 1990, ITS-90. A temperature in the unit °C of this scale is denoted by the symbol t, except in Table A.1 where the full nomenclature t90 /°C is used. This document applies to platinum resistors whose temperature coefficient a, defined as a= R100 - R0 R0 · 100°C is conventionally written as a = 3,851·10-3 °C-1, where R100 is the resistance at t = 100 °C and R0 is the resistance at t = 0 °C. This document covers platinum resistors and thermometers for the temperature range -200 °C to +850 °C with different tolerance classes. It can also cover particular platinum resistors or thermometers for a part

50.62

Status : Valid
Approval date : 4/29/2022
Number of pages : 56
ICS : 17.200.20 Temperature measuring instruments
Technical Committee : 23 - Industrial -process measurement and control

Relations with other standards: