SR EN IEC 61558-2-15:2025

Safety of transformers, reactors, power supply units and combinations thereof - Part 2-15: Particular requirements and tests for isolating transformers for medical IT systems for the supply of medical locations

This part of IEC 61558 deals with safety of isolating transformers for medical IT systems for the supply of medical locations. NOTE 1 Safety includes electrical, thermal and mechanical aspects. Unless otherwise specified, from here onward, the term transformer covers isolating transformers for medical IT systems for the supply of medical locations. This document is applicable to isolating transformers, single-phase or three-phase, air-cooled (natural or forced) dry-type transformers for the supply of medical IT system for group 2 medical locations, designed to be permanently connected to the fixed wiring and intended to form the medical IT system on the secondary side. The windings can be encapsulated or non encapsulated. NOTE 2 IT systems are defined in IEC 60364-1. The installation rules for medical IT system for group 2 medical locations are covered by IEC 60364-7-710. Transformers covered by this document are intended for medical IT systems for the supply of medical locations. All other transformers or equipments connected downstream from the transformer are not covered by this

45.30

Status : Valid
Approval date : 4/30/2025
Number of pages : 50
ICS : 29.180 Transformers. Reactors
Technical Committee : 312 - Small power transformers

Relations with other standards: